Description
High-precision 12-lead synchronous acquisition
MK-1206A captures all 12 leads simultaneously and runs a built-in ECG interpretation algorithm for reliable cardiac screening and early arrhythmia assessment. Defibrillation protection and high CMRR suppress muscle and power-line interference, keeping signals clean even in demanding clinical settings.
Intuitive touch-control & efficient workflow
The 7-inch HD color touchscreen shows waveforms, patient data, and device status at a glance; touch entry and direct report printing keep work moving. Auto/manual/rhythm modes, real-time lead-off alerts, and large local storage for review and re-print make clinic and bedside exams faster and smoother.
Features
- 7-inch high-definition color touch LCD screen, 800×480 resolution for real-time ECG waveform display
- 12-lead synchronous ECG acquisition, 6-channel thermal print output
- Built-in auto ECG interpretation algorithm, supports auto / manual / rhythm-analysis working modes
- Defibrillation-protected floating input circuit, high anti-interference performance
- Lead-off detection & alarm function, real-time prompt for poor electrode contact
- Large-capacity local case storage, support reviewing and re-printing historical reports
- Thermal printing system, compatible with 110mm width thermal recording paper
- AC-DC dual-power supply with rechargeable lithium-ion battery, suitable for mobile inspection
- Compact portable chassis, easy for deployment among different clinical locations
- Meet IEC 60601-1 medical electrical safety standard, CF-type patient isolation safety grade
Specifications
Display
| Item | Parameter |
|---|
| Screen | 7-inch color touch LCD |
| Resolution | 800 × 480 pixels |
ECG Acquisition
| Item | Parameter |
|---|
| Leads | Standard 12-lead, 12-lead simultaneous acquisition |
| Sampling Rate | 8000 samples / second |
| Frequency Response | 0.05 Hz–150 Hz |
| Time Constant | ≥3.2 s |
| CMRR | >60dB; >100dB with filter enabled |
| Calibration Voltage | 1 mV ±5% |
| Input Dynamic Range | ±500 mV |
| Anti-interference Filter | 50/60Hz AC filter, multi-grade EMG filter, baseline drift filter |
Printing & Storage
| Item | Parameter |
|---|
| Print Mode | 6-channel / 6+1-channel thermal printing |
| Paper Spec. | 110mm width thermal roll paper |
| Paper Speed | 5 / 10 / 25 / 50 mm/s |
| Case Storage | Large-capacity local memory for patient ECG records, support review & re-print |
Environmental
| Item | Parameter |
|---|
| Operating Temperature | 5 ℃–40 ℃ |
| Operating Humidity | 25%–95% non-condensing |
| Atmospheric Pressure | 700–1060 hPa |
Power Requirement
| Item | Parameter |
|---|
| AC Input | 110–230V~, 50/60Hz |
| Battery | Built-in rechargeable lithium-ion battery, support continuous battery-driven operation |
Mechanical
| Item | Parameter |
|---|
| Weight | 3.5kg (gross weight with accessories) |
| Safety Class | Class I, Type CF patient isolation, IEC60601-1 compliant |
Clinical Applications
Built for hospitals, clinics, community health centers, physical examination centers, and pre-hospital emergency settings, the MK-1206A supports routine screening, bedside acquisition, high-volume checkups, and rapid on-scene ECG capture—wherever reliable cardiac data is needed.
Product Note: This is a Class II medical device (Guangdong Medical Device Registration No. 20202070169). It is intended for medical institutions to acquire human ECG waveforms for clinical diagnosis and research. It must not be used as the sole basis for a confirmed diagnosis; the final diagnosis should be made by a licensed physician based on comprehensive evaluation.
Comprehensive score