0

ECG-5512G Device

Quantity:
Contact Now
add to cart
Buy Now
ECG-5512G
Country of Origin: 
Guangzhou

Description

 
 

All Twelve Leads, in a Single Acquisition

Every standard lead is acquired simultaneously at 4,000 Hz per channel, so one acquisition delivers a complete resting ECG. Review the full twelve-lead waveform on screen before using any paper, and follow either Wilson or Cabera lead sequence.

MAN, AUTO, R-R and STOR modes let the operator match the acquisition to the exam: routine printed report, rhythm strip or paper-free record.

Signal Integrity Where It Matters

With the AC filter engaged, common-mode rejection reaches ≥120 dB, keeping mains interference out of the trace. ADS baseline-drift suppression works alongside selectable AC (50/60 Hz) and EMG (25/45 Hz) filters.

A floating input stage with ≥50 MΩ input impedance and system noise below 30 μVp-p helps preserve the fine detail that matters. 

 
 

Touch Control, Thermal Print Output

A 9-inch 800 × 480 colour touch LCD and 11 hard keys put patient data, filters and recording format at your fingertips—set them on screen or from the keypad. Four interface themes and five backlight levels adapt the display to different rooms and working distances.

Recording is handled by a thermal dot-matrix printer on 210 mm × 30 m roll paper, with 3ch+++, 6ch, 6ch+ or 12ch layouts, 6.25 to 50 mm/s paper speed and 2.5 to 40 mm/mV sensitivity.

Cases That Stay with the Department

Keep up to 400 patient cases in internal memory—and add an SD card when you need more room. Move files to a USB drive or to PC ECG management software through the USB 2.0 host port, and print stored reports on plain A4 paper with a supported HP PCL5 printer.

A barcode scanner connected to the UART port enters patient numbers in a single scan, so a screening list and its recordings stay matched.

 

 

Features

  • Twelve-lead synchronous acquisition at 4,000 Hz per channel, with pacing-pulse detection.
  • Four working modes: MAN, AUTO (review or real-time), R-R rhythm and STOR.
  • Recording layouts 3ch+++, 6ch, 6ch+ and 12ch, with short, normal, long and long+ auto length.
  • 9-inch 800 × 480 colour touch LCD with 11 keys, four themes and five backlight levels.
  • ADS baseline-drift filter plus AC 50/60 Hz and EMG 25/45 Hz filters; CMRR ≥ 120 dB.
  • Defibrillation-protected CF applied part with floating input; patient leakage current < 10 μA.
  • Thermal dot-matrix recorder on 210 mm × 30 m roll paper, 6.25 to 50 mm/s, 2.5 to 40 mm/mV.
  • 400 patient cases on board, SD card optional, with USB export to drive or PC software.
  • USB link to HP PCL5 A4 printers; UART port for barcode-based patient ID entry.
  • 3.0 kg 310 × 278 × 76 mm with carry handle; 4,400 mAh battery for 8 hours of cordless use.

Specifications

Display

LCD9-inch colour touch LCD, 800 × 480 graphic
BacklightFive brightness levels
Colour themesFour interface themes

Keypad and Lead View

Keypad11 keys, plus touch input
Lead viewSimultaneous twelve-lead waveform display

ECG Acquisition

LeadsStandard twelve leads, synchronous acquisition
Input circuitFloating, protection against defibrillator effect
Input impedance≥50 MΩ
Electrode DC current≤0.1 μA
Input dynamic range±5 mVp-p, DC bias ≥ ±620 mV
CMRR≥120 dB (with AC filter)
Patient leakage current<10 μA
Calibration voltage1 mV ±5%
Sampling rate4,000 Hz per channel
Heart rate range30–300 bpm (±1 bpm)
Sensitivity2.5 / 5 / 10 / 20 / 40 mm/mV (±3%), Auto
Paper speed6.25 / 12.5 / 25 / 50 mm/s (±3%)
System error< ±5% or < ±40 μV
System noise<30 μVp-p
Inter-channel interference<2%
Lag<0.5 mm
Pacing pulse displayAmplitude 2–250 mV, width 0.1–2.0 ms

Filters

AC filter50 Hz / 60 Hz / off
EMG filter25 Hz / 45 Hz / off
Baseline driftADS filter

Recorder

Recording methodThermal dot-matrix
Recording format3ch+++, 6ch, 6ch+, 12ch
Paper210 mm × 30 m, roll thermal paper

Storage & Connectivity

Patient cases400 cases in internal memory, SD card optional
USB hostUSB 2.0, FAT32 USB drive
UART115 kbps, 3.3 V level; barcode scanner and factory service
External printingUSB printer, HP PCL5
PC softwareECG management software over USB

Power

AC input100–240 V (±10%), 50/60 Hz (±1 Hz), 60 VA
Battery14.4 V / 4,400 mAh lithium-ion
Charging current1,300 mA, constant current / constant voltage
Charge cyclesNot less than 500
Battery operation8 hours from a full charge
FuseTZAH/250 W, 5 × 20 mm

Physical & Environmental

Dimensions310 × 278 × 76 mm
Weight3.0 kg
Operating temperature+5 °C to +40 °C
Operating humidity25%–80% RH, non-condensing
Operating pressure700–1,060 hPa
Storage / transport temperature-20 °C to +55 °C
Storage / transport humidity25%–93% RH

Safety & Compliance

Protection classClass I, internally powered equipment
Applied partType CF, defibrillation-proof
StandardsIEC 60601-1, IEC 60601-2-25, IEC 60601-1-2
EMCCISPR 11 Group 1, Class A
Liquid protectionOrdinary equipment, no water protection

Clinical Applications

From routine outpatient ECG rooms and bedside emergencies to paediatric cardiology, community screening and clinical teaching, the ECG-5512G adapts to the setting at hand.