VS2000 NIBP Calibration Guide
VS2000 NIBP Calibration Guide
1. Prepare the VS2000 machine and sphygmomanometer, the connection method is as follows:


2. Select the menu and enter the “Maintenance” option;



3. Select the “Login” option, Type is dealer, enter the account password to log in, the password is "0910";


4.Select the "Calibrate" option;

5.Select the "NIBP" option;


6.Select Level: 0mmHg for calibration,the steps are as follows:



Step 1: Click "Start" to calibrate.



Step 2: The calibration is successful, click "Stop" to end the calibration.



Step 3: Click "Read" to read the information.



Step 4: Complete the calibration.
7.Select Level: 200mmHg for calibration, the steps are as follows:




Step 1: Click "Start" to perform calibration.



Step 2: Pressurize VS2000, the pressure value is 200mmHg.



Step 3: After the calibration is successful, click "Stop" to end the calibration.



Step 4: Click "read" to read the information.



Step 5: Click "OK" to complete the calibration.