VS2000 NIBP Calibration Guide
1. Prepare the VS2000 machine and sphygmomanometer, the connection method is as follows:
2. Select the menu and enter the “Maintenance” option;
3. Select the “Login” option, Type is dealer, enter the account password to log in, the password is "0910";
4.Select the "Calibrate" option;
5.Select the "NIBP" option;
6.Select Level: 0mmHg for calibration,the steps are as follows:
Step 1: Click "Start" to calibrate.
Step 2: The calibration is successful, click "Stop" to end the calibration.
Step 3: Click "Read" to read the information.
Step 4: Complete the calibration.
7.Select Level: 200mmHg for calibration, the steps are as follows:
Step 1: Click "Start" to perform calibration.
Step 2: Pressurize VS2000, the pressure value is 200mmHg.
Step 3: After the calibration is successful, click "Stop" to end the calibration.
Step 4: Click "read" to read the information.
Step 5: Click "OK" to complete the calibration.