VS2000 NIBP Calibration Guide

author: Zhoupeng
23/03/2022

VS2000 NIBP Calibration Guide

1. Prepare the VS2000 machine and sphygmomanometer, the connection method is as follows:

9

2. Select the menu and enter the “Maintenance” option;

8

3. Select the “Login” option, Type is dealer, enter the account password to log in, the password is "0910";


4.Select the "Calibrate" option;

5.Select the "NIBP" option;

6.Select Level: 0mmHg for calibration,the steps are as follows:

2

Step 1: Click "Start" to calibrate.

1

Step 2: The calibration is successful, click "Stop" to end the calibration.

17

Step 3: Click "Read" to read the information.

15

Step 4: Complete the calibration.
 

7.Select Level: 200mmHg for calibration, the steps are as follows:

13

Step 1: Click "Start" to perform calibration.

11

Step 2: Pressurize VS2000, the pressure value is 200mmHg.

18

Step 3: After the calibration is successful, click "Stop" to end the calibration.

20

Step 4: Click "read" to read the information.

19

Step 5: Click "OK" to complete the calibration.