UT-120V Veterinary Blood and Infusion Warmer
Description

Precise warming, dual safety
UT-120V stabilizes blood and infusion fluids at 32–42 °C with an accuracy of ±1 °C. Dual overheat protection via hardware and software activates at 50 ± 5 °C, delivering double safeguards for patients and equipment during prolonged surgical procedures.
Dual channels, independent control
Dual channels operate synchronously, with each channel featuring independently set temperature, real-time heating temperature, heating duration and alarm status. All information is clearly displayed side by side on the 72 × 72 mm screen, suitable for surgical scenarios requiring multi-channel infusion or blood transfusion.


Flexible dry silicone, vet-ready
Five types of detachable medical-grade silicone profiles (60 / 90 / 100 / 120 / 140 cm) are compatible with 3.5 / 5 / 7 mm PVC blood transfusion/infusion tubing. Dry heat transfer eliminates the risk of fluid wetting, with preheating completed within 4 minutes.
The compact unit measures 155 × 110 × 170 mm and weighs 1.2 kg. It supports continuous heating mode, features IPX4 protection and Class II Type CF electric shock protection, and fits surgical trays, ICU stands and emergency trolleys. A comprehensive alarm system enables clinical staff to monitor the status throughout the process.
Features
Specifications
Temperature Control
Item | Value |
|---|---|
Temperature Setting | 32.0–42.0 °C (86.9–107.6 °F) |
Temperature Deviation | ≤ ±1 °C (≤ ±1.8 °F) |
Temperature Increment | 0.1 °C (0.2 °F) |
Optional Unit | °C / °F |
Warming Up Time | < 4 min |
Overheat Protection | 50 ± 5 °C, hardware and software protection |
Alarm | Overheat, low temperature, system error, over-time, warming up |
Mute Time | 2 min |
Heating System
Item | Value |
|---|---|
Heating Way | Dry silicone heating profile |
Heating Profiles | 60 / 90 / 100 / 120 / 140 cm detachable, optional |
Applicable Blood/IV Sets | Standard single-use PVC blood/IV tubes, diameter 3.5–7 mm (3.5 / 5 / 7 mm) |
Heating Time Range | 00h00min–99h59min |
Working Mode | Continuous heating |
Display
Item | Value |
|---|---|
Display Size | 72 × 72 mm |
Display Content | Set temperature, heating temperature, heating time, alarm message, working status (dual channel) |
Power
Item | Value |
|---|---|
Power Supply | 100–240 V~ 50/60 Hz |
Power Consumption | 80 VA |
Physical & Protection
Item | Value |
|---|---|
Dimension | 155 × 110 × 170 mm |
Weight | 1.2 kg (including one 120 cm heating profile) |
Degree of Protection against Liquids | IPX4 |
Degree of Protection against Electric Shock | Class II, Type CF |
Environment
Item | Value |
|---|---|
Operation Environment | +5 °C to +30 °C; RH 20%–93%; Atmosphere 70.0–106.0 kPa |
Transportation and Storage | -20 °C to +55 °C; RH 10%–95%; Atmosphere 50.0–106.0 kPa |
Clinical Applications

Comprehensive score