The monitor plays an important role in the entire monitoring process. Since the monitor works almost 24 hours a day, its failure rate is also high. The common faults and troubleshooting methods are introduced as follows:

Fault phenomenon:
When the instrument is turned on, the screen does not display, and the indicator light does not light; when the external power supply is connected, the battery voltage is low, and the machine automatically shuts down; useless.
Inspection Method:
① When the instrument is not connected to AC power, check whether the 12V voltage is low. This fault alarm indicates that the output voltage detection part of the power supply board detects that the voltage is low, which may be caused by the failure of the power supply board detection part or the output failure of the power supply board, or it may be caused by the failure of the back-end load circuit.
②When there is a battery installed, this phenomenon means that the monitor is working on battery power supply and the battery power is basically used up, and the AC input does not work normally. The possible reasons are: the 220V power socket itself has no power, or the fuse is blown.
③ When no external battery is connected, it may be judged that the rechargeable battery is broken, or the battery cannot be charged due to the failure of the power board/charging control board.
Remedy: Connect all the connection parts reliably, and connect the AC power to charge the instrument.
The monitor plays an important role in the entire monitoring process. Since the monitor works almost 24 hours a day, its failure rate is also high. The common faults and troubleshooting methods are introduced as follows:

Fault phenomenon:
When the instrument is turned on, the screen does not display, and the indicator light does not light; when the external power supply is connected, the battery voltage is low, and the machine automatically shuts down; useless.
Inspection Method:
① When the instrument is not connected to AC power, check whether the 12V voltage is low. This fault alarm indicates that the output voltage detection part of the power supply board detects that the voltage is low, which may be caused by the failure of the power supply board detection part or the output failure of the power supply board, or it may be caused by the failure of the back-end load circuit.
②When there is a battery installed, this phenomenon means that the monitor is working on battery power supply and the battery power is basically used up, and the AC input does not work normally. The possible reasons are: the 220V power socket itself has no power, or the fuse is blown.
③ When no external battery is connected, it may be judged that the rechargeable battery is broken, or the battery cannot be charged due to the failure of the power board/charging control board.
Remedy: Connect all the connection parts reliably, and connect the AC power to charge the instrument.
Related News
What is noninvasive blood pressure (NIBP)?
2023-09-28 7Clinical Application and Significance of End-Tidal Carbon Dioxide(ETCO2) Monitoring
2023-09-27 8Popular Science|How to choose home medical devices
2023-09-26 14Pulse oximeter can help you check your body
2023-05-09 408How do pulse oximeters work?
2023-04-03 557Why is the SpO2 reading on the pulse oximeter fluctuate between 92 and 97? Is this normal?
2023-03-30 577Different measurement positions lead to different measurement results
2023-03-24 453Blood Oximeter: A New Concept Family Doctor
2023-03-23 436Significance of monitoring BIS for anesthetized patients
2023-02-21 537Which gases need to be detected when the patient is anesthetized
2023-02-20 514What are the types of disposable spo2 probe?
2023-01-06 612UM2012 Patient Monitor has successfully passed the EMC test
2022-11-07 71115 products from UTECH Medical obtained FSC issued by the European Union
2022-11-04 669UTECH participated in the 14th East-West Small Animal Clinical Veterinary Exhibition
2022-11-02 565UTECH live room Highlights
2022-10-31 536Brazil' INMETRO audited UTECH's new oximeter uPM50
2022-10-28 487Patient Monitor UM2012 passed YY0505 testing
2022-10-26 439Nation's capacity in treatment of heart disease doesn't skip a beat
2022-09-27 474Increasing trend of remote-monitoring is boosting the market growth
2022-09-26 388CDSCO – Medical Device Registration in India
2022-09-22 311