IP300 VET INFUSION PUMP

IP300 VET INFUSION PUMP
IP300V
Changsha
This veterinary infusion pump consists of cabinet, control panel, display and optional drop sensor. It is operated by trained staff in medical institutions to deliver medicines and nutrients with precise flow‑rate control. It is not for home‑care or enteral feeding, and blood products infusion is prohibited.

It supports rate, time, dose, sequential and drug‑library infusion modes, along with bolus, purge and KVO functions. Equipped with audio‑visual alarms and built‑in rechargeable battery, it also offers optional warmer and WiFi modules. Its flow rate ranges from 0.1‑1500 ml/h with ±4.5% accuracy within 12 hours. Rated IPX4 for splash‑proof, the pump has a 7‑year service life. Operators shall respond promptly to device alarms.
Quantity

Description

Precision infusion, covering flow rates across all application scenarios

The IP300V veterinary infusion pump offers an adjustable infusion flow rate ranging from 0.1 to 1,500 ml/h, with an infusion accuracy of ±4.5% (over 12 hours, compliant with EN/IEC 60601‑2‑24), meeting diverse infusion requirements—from low‑rate maintenance to rapid fluid replacement.

It features built‑in drip rate detection and 14‑level blockage pressure thresholds, ensuring stable and controllable infusion accuracy for clinical applications.

                                     

 

Multiple infusion modes for flexible adaptation

The system supports infusion modes including infusion rate, duration, dose, sequence (optional), and drug library (optional); these modes can be switched with a single press of the"Mode"button.

The dose mode allows dose calculation based on animal body weight (0.1–300 kg), making it suitable for animals of varying boy sizes.
 

 
    

 

Multi‑layered safety protection for peace of mind during infusion

With pipeline bubble detection, real‑time blockage monitoring, intelligent tiered alerting, and anti‑push‑in functionality—combined with the KVO intravenous access mode—this system ensures full‑process infusion safety.

 
                          

 
 

Portable and durable, featuring connectivity and heating capabilities

Equipped with a built‑in lithium battery (providing approximately 5 hours of runtime on a full charge), enabling continuous infusion during power outages or transportation.

Optional tubing heater and WiFi connectivity support facilitate deployment in mobile scenarios and centralized management.

Features

  • Multi‑infusion mode: Rate / Time / Dose / Sequence (optional) / Drug Library (optional); switch modes with a single press of the Mode key.
  • Manual and automatic bolus dosing functions enable rapid drug administration.
  • Purge function: rapidly discharge air from the piping system.
  • For the KVO intravenous open mode, after completion of infusion, maintain patency of the access line at a rate of 0.1–20 ml/h.
  • The infusion flow rate can be adjusted in real time to meet clinical needs.
  • Drop rate sensor (optional): automatically detects and counts drops to monitor the infusion rate.
  • Pipeline heater (optional, 30–50 °C) warms the infusion fluid to reduce irritation to the animal.
  • WiFi (optional, 2.4 GHz) uploads the operating status of the device to the infusion monitoring system.
  • Intelligent tiered alerts: bubbles, blockage, absence of medicinal fluid, low battery level, free flow, etc.; alerts can be silenced or cleared.
  • 10‑level brightness adjustment, multilingual interface, and support for storing and exporting up to 5,000 historical records.

Specifications

Infusion performance
Parameter itemsparameter values
Infusion flow rate0.1–1500 ml/h (step sizes: 0.01 / 0.1 / 1 ml/h)
Drop rate setting1–400 drops/min
Infusion accuracy±4.5% (12 h, 1 L infusion, compliant with EN/IEC 60601‑2‑24)
Intravenous fluid volume (VTBI)0.0~9999.9 ml
Infusion duration00:00:01~99:59:59(hh:mm:ss)
Piping flow rate0.1–1500 ml/h (default: 600 ml/h)
Infusion rate0.1~1500 ml/h
Push volume1.0~100.0 ml
KVO speed0.1~20 ml/h
Dose regimen
Parameter itemsparameter values
Volume of medicinal solution0.1–999.9 (Units: g / mg / U / IU)
dosage0.001~9999
Animal body weight0.1~300.0 kg
Quality of the medicinal solution0.1~9999.9 ml
Alarm and Safety
Parameter itemsparameter values
Obstruction pressure rating14 pressure stages (30–160 kPa, with an error of ±15 kPa per stage)
Maximum blocking pressure200 kPa
Bubble DetectionSingle bubble: 15–1000 μl; cumulative volume: 0.1–2 ml per 15 min
Power source
Parameter itemsparameter values
alternating current power supply220 V, 50 Hz, with ground fault protection
fuseF2AL/250 V ×2
maximum power dissipation60 VA
power line2.5 m
Battery
Parameter itemsparameter values
typeLithium polymer battery, 7.2 Vdc, 2600 mAh (approx. 120 g)
duration of flightApproximately 5 hours (at flow rates of 25 ml/h and below); approximately 4 hours at maximum flow rate
charging intervalPower‑off duration <10 h; Power‑on duration <14 h
Dimensions and Weight
Parameter itemsparameter values
main engine120 × 100 × 215 mm (excluding column clamps), approximately 1.5 kg
pack248 × 190 × 155 mm, approximately 2.1 kg
ambient condition
Parameter itemsparameter values
working temperature5~40 ℃
Working humidity20%–80% (no coagulation)
atmospheric pressure70~106 kPa
Standards and Protection
Parameter itemsparameter values
safety standardsIEC 60601‑1 / 60601‑2‑24 / 60601‑1‑2 / 60601‑1‑8 / IEC 60529
levels of protectionIPX4 (Spill‑proof)
Application ComponentsDefibrillation CF‑type application component
Electrical TypeClass I devices
Installation column diameter15~36 mm

Clinical Applications

      

                      hospital                                            Animal Clinic                                      Operating Room