It supports rate, time, dose, sequential and drug‑library infusion modes, along with bolus, purge and KVO functions. Equipped with audio‑visual alarms and built‑in rechargeable battery, it also offers optional warmer and WiFi modules. Its flow rate ranges from 0.1‑1500 ml/h with ±4.5% accuracy within 12 hours. Rated IPX4 for splash‑proof, the pump has a 7‑year service life. Operators shall respond promptly to device alarms.
Description
Precision infusion, covering flow rates across all application scenarios
The IP300V veterinary infusion pump offers an adjustable infusion flow rate ranging from 0.1 to 1,500 ml/h, with an infusion accuracy of ±4.5% (over 12 hours, compliant with EN/IEC 60601‑2‑24), meeting diverse infusion requirements—from low‑rate maintenance to rapid fluid replacement.
It features built‑in drip rate detection and 14‑level blockage pressure thresholds, ensuring stable and controllable infusion accuracy for clinical applications.

Multiple infusion modes for flexible adaptation
The system supports infusion modes including infusion rate, duration, dose, sequence (optional), and drug library (optional); these modes can be switched with a single press of the"Mode"button.
The dose mode allows dose calculation based on animal body weight (0.1–300 kg), making it suitable for animals of varying boy sizes.

Multi‑layered safety protection for peace of mind during infusion
With pipeline bubble detection, real‑time blockage monitoring, intelligent tiered alerting, and anti‑push‑in functionality—combined with the KVO intravenous access mode—this system ensures full‑process infusion safety.

Portable and durable, featuring connectivity and heating capabilities
Equipped with a built‑in lithium battery (providing approximately 5 hours of runtime on a full charge), enabling continuous infusion during power outages or transportation.
Optional tubing heater and WiFi connectivity support facilitate deployment in mobile scenarios and centralized management.
Features
- Multi‑infusion mode: Rate / Time / Dose / Sequence (optional) / Drug Library (optional); switch modes with a single press of the Mode key.
- Manual and automatic bolus dosing functions enable rapid drug administration.
- Purge function: rapidly discharge air from the piping system.
- For the KVO intravenous open mode, after completion of infusion, maintain patency of the access line at a rate of 0.1–20 ml/h.
- The infusion flow rate can be adjusted in real time to meet clinical needs.
- Drop rate sensor (optional): automatically detects and counts drops to monitor the infusion rate.
- Pipeline heater (optional, 30–50 °C) warms the infusion fluid to reduce irritation to the animal.
- WiFi (optional, 2.4 GHz) uploads the operating status of the device to the infusion monitoring system.
- Intelligent tiered alerts: bubbles, blockage, absence of medicinal fluid, low battery level, free flow, etc.; alerts can be silenced or cleared.
- 10‑level brightness adjustment, multilingual interface, and support for storing and exporting up to 5,000 historical records.
Specifications
| Parameter items | parameter values |
|---|---|
| Infusion flow rate | 0.1–1500 ml/h (step sizes: 0.01 / 0.1 / 1 ml/h) |
| Drop rate setting | 1–400 drops/min |
| Infusion accuracy | ±4.5% (12 h, 1 L infusion, compliant with EN/IEC 60601‑2‑24) |
| Intravenous fluid volume (VTBI) | 0.0~9999.9 ml |
| Infusion duration | 00:00:01~99:59:59(hh:mm:ss) |
| Piping flow rate | 0.1–1500 ml/h (default: 600 ml/h) |
| Infusion rate | 0.1~1500 ml/h |
| Push volume | 1.0~100.0 ml |
| KVO speed | 0.1~20 ml/h |
| Parameter items | parameter values |
|---|---|
| Volume of medicinal solution | 0.1–999.9 (Units: g / mg / U / IU) |
| dosage | 0.001~9999 |
| Animal body weight | 0.1~300.0 kg |
| Quality of the medicinal solution | 0.1~9999.9 ml |
| Parameter items | parameter values |
|---|---|
| Obstruction pressure rating | 14 pressure stages (30–160 kPa, with an error of ±15 kPa per stage) |
| Maximum blocking pressure | 200 kPa |
| Bubble Detection | Single bubble: 15–1000 μl; cumulative volume: 0.1–2 ml per 15 min |
| Parameter items | parameter values |
|---|---|
| alternating current power supply | 220 V, 50 Hz, with ground fault protection |
| fuse | F2AL/250 V ×2 |
| maximum power dissipation | 60 VA |
| power line | 2.5 m |
| Parameter items | parameter values |
|---|---|
| type | Lithium polymer battery, 7.2 Vdc, 2600 mAh (approx. 120 g) |
| duration of flight | Approximately 5 hours (at flow rates of 25 ml/h and below); approximately 4 hours at maximum flow rate |
| charging interval | Power‑off duration <10 h; Power‑on duration <14 h |
| Parameter items | parameter values |
|---|---|
| main engine | 120 × 100 × 215 mm (excluding column clamps), approximately 1.5 kg |
| pack | 248 × 190 × 155 mm, approximately 2.1 kg |
| Parameter items | parameter values |
|---|---|
| working temperature | 5~40 ℃ |
| Working humidity | 20%–80% (no coagulation) |
| atmospheric pressure | 70~106 kPa |
| Parameter items | parameter values |
|---|---|
| safety standards | IEC 60601‑1 / 60601‑2‑24 / 60601‑1‑2 / 60601‑1‑8 / IEC 60529 |
| levels of protection | IPX4 (Spill‑proof) |
| Application Components | Defibrillation CF‑type application component |
| Electrical Type | Class I devices |
| Installation column diameter | 15~36 mm |
Clinical Applications

/5