VS2000 NIBP Calibration Guide
VS2000 NIBP Calibration Guide
VS2000 NIBP Calibration Guide
1. Prepare the VS2000 machine and sphygmomanometer, the connection method is as follows:
2. Select the menu and enter the “Maintenance” option;
3. Select the “Login” option, Type is dealer, enter the account password to log in, the password is "0910";
4.Select the "Calibrate" option;
5.Select the "NIBP" option;
6.Select Level: 0mmHg for calibration,the steps are as follows:
Step 1: Click "Start" to calibrate.
Step 2: The calibration is successful, click "Stop" to end the calibration.
Step 3: Click "Read" to read the information.
Step 4: Complete the calibration.
7.Select Level: 200mmHg for calibration, the steps are as follows:
Step 1: Click "Start" to perform calibration.
Step 2: Pressurize VS2000, the pressure value is 200mmHg.
Step 3: After the calibration is successful, click "Stop" to end the calibration.
Step 4: Click "read" to read the information.
Step 5: Click "OK" to complete the calibration.
Related News
Clinical application scenarios of the patient monitor?
2023-02-16 424How does blood oxygen hypoxia do?
2022-12-14 521The correct use of SpO2 sensor method?
2022-12-13 552Theory of measuring human blood pressure
2022-11-10 515What is a hyperbaric oxygen chamber?
2022-10-27 474What causes an irregular heartbeat?
2022-10-25 481What reason is human body temperature low?
2022-10-24 442Who is oxygen therapy suitable for?
2022-10-20 460How should blood oxygen do low?
2022-10-19 433What is the normal range of respiratory rate?
2022-10-17 335Is newborn's body temperature higher than adult's?
2022-10-14 367Which cardiac tests are appropriate for the heart?
2022-10-13 303Why should patients fast for 6 hours after anesthesia?
2022-10-12 269What should we do if the blood oxygen saturation is low?
2022-10-11 308Newborn blood oxygen saturation
2022-10-10 294Ear Pulse Oximeters
2022-10-09 237Introduction to the key points of Patient Monitor operation
2022-10-08 241Relationship between blood oxygen saturation and heart rate
2022-10-02 258What does SPO2 mean on the patient monitor
2022-10-01 253What are the indications of the patient monitor
2022-10-01 1268