Basis of Respiratory Parameters on the Monitor
2022-06-23
Basis of Respiratory Parameters on the Monitor
Principle of respiration measurement
Due to the breathing activity of the patient's thorax, the impedance changes between the RA and LL electrodes. The monitor measures the respiration rate by detecting the change in the thoracic impedance between the RA and LL electrodes, and generates a breathing wave on the screen.

Factors Affecting Respiratory Monitoring
To monitor breathing, no additional electrodes are required, but the placement of electrodes is very important. Some patients, due to their clinical condition, laterally expand their thorax resulting in negative intrathoracic pressure. In this case, it is best to place the two respiration electrodes on the right midaxillary line and the left thoracic area where breathing is most active to obtain optimal respiration waves.
Second: Place the white RA (right arm) and red LL (right leg) electrodes diagonally for optimal respiration waves. The hepatic region and ventricle should be avoided in line with the respiratory electrodes, so as to avoid cardiac overlay or artefacts from pulsatile blood flow, which is especially important in neonates.
Third: Respiratory monitoring is not suitable for patients with high range of motion, as this may lead to false alarms.
Principle of respiration measurement
Due to the breathing activity of the patient's thorax, the impedance changes between the RA and LL electrodes. The monitor measures the respiration rate by detecting the change in the thoracic impedance between the RA and LL electrodes, and generates a breathing wave on the screen.

Factors Affecting Respiratory Monitoring
To monitor breathing, no additional electrodes are required, but the placement of electrodes is very important. Some patients, due to their clinical condition, laterally expand their thorax resulting in negative intrathoracic pressure. In this case, it is best to place the two respiration electrodes on the right midaxillary line and the left thoracic area where breathing is most active to obtain optimal respiration waves.
Second: Place the white RA (right arm) and red LL (right leg) electrodes diagonally for optimal respiration waves. The hepatic region and ventricle should be avoided in line with the respiratory electrodes, so as to avoid cardiac overlay or artefacts from pulsatile blood flow, which is especially important in neonates.
Third: Respiratory monitoring is not suitable for patients with high range of motion, as this may lead to false alarms.
Common Causes of errors in blood pressure measurement
The operation of the patient monitor
Related Article
Discover how summer heat affects the body and learn the difference between being overheated and having a fever. This guide provides practical tips on how to correctly measure body temperature in hot weather and highlights the features of UTECH Medical’s UTM10 infrared thermometer, including its one-second, non-contact measurement, 30-reading memory, and reliable accuracy for home and public safety.
Too Hot? Check Your Temperature the Right Way This Summer
UTECH Medical conducts customs declaration and Certificate of Origin (COO) training to strengthen employees' export knowledge and provide global customers with more efficient medical equipment transaction support.
UTMI Training Strengthens Customs Expertise and Customer Service

Learn key facts about Marburg virus disease (MVD), a rare but severe hemorrhagic fever caused by a Filoviridae RNA virus. Discover how the virus spreads from Egyptian rousette bats to humans, its initial symptoms—including high fever, bleeding, and organ failure—and current laboratory diagnosis methods like RT-PCR. Explore supportive care treatments, prevention strategies
Marburg virus disease